Choosing a Detector

Pick what you need to measure, see which detector type suits it and what we carry for it.

Knowledge Base

What do you need to measure?

Detector types compared

The table gives relative ranking, not device figures: two instruments built on the same detector type can differ markedly. For exact energy resolution, efficiency and measuring range, consult the datasheet of the instrument in question.

DetectorEnergy resolutionEfficiencyIdentifies nuclidesRobustnessTypical use
Geiger-Müller counternonelowvery highDose rate measurement, leak searching, wide-range monitoring
Plastic scintillatornonevery highhighPortal monitor, vehicle and baggage screening, large-volume scanning
NaI(Tl) scintillatormediumhighmediumField spectrometry, nuclide identification, environmental monitoring
LaBr₃(Ce) / SrI₂(Eu) scintillatorgoodhighmediumPrecise nuclide identification in the field, separating mixed sources
CdZnTe (CZT) semiconductorgoodlowhighSmall handheld spectrometer, low-energy gamma and X-rays
HPGe semiconductorvery goodmediumlowLaboratory gamma spectrometry, quantitative analysis, sample measurement
He-3 / proportional counternonemediumhighNeutron measurement, fissile material detection, portal neutron channel
Wide-window proportional counter / ZnS(Ag)nonehighmediumSurface contamination, alpha/beta contamination checks
Silicon semiconductormediumlowhighElectronic personal dosimeter, accumulated dose tracking

The number of filled dots shows the relative level: one dot low, four very high. An empty row means the detector does not offer that property at all (a Geiger-Müller counter, for instance, does not discriminate energy).

This comparison is deliberately qualitative. Energy resolution and efficiency vary with crystal volume, temperature, electronics and measuring geometry even within one detector type. Rather than publish individual figures, relative ranking is given; where a number matters, the instrument's own datasheet governs. The single numeric example in the table (3.2% at 662 keV for SrI₂(Eu)) comes from the manufacturer's product page.